8.1 Error Messages and Error Handling
KP3000GS
207
8.1.3 If An Error Occurs Repeatedly
If an error message repeatedly appears even though you removed the cause of the error, the error
may be resolved by restarting the power unit.
If the error still exists, this product or the power unit in the product may be faulty. Repair is needed
in this case. Contact the seller (NF Corporation or our agent) from whom you purchased the product.
Even though one of the power units is faulty, the product can be used by disconnecting the faulty
power unit and using the rest of them.
The following shows how to restart and disconnect the power unit in the case of an error.
--------
Notes
------------------------------------------------------------------------------------------------------------
The following procedure is available only when the [PU-ON] soft-key is displayed on the error
message window. Stop using this product in the case of a repeating error without the [PU-ON]
soft-key.
-----------------------------------------------------------------------------------------------------------------------------
Operation procedure
1.
In the error message window, press the [PU-ON] soft-key. The power unit energization setting
window opens.
2.
The power unit with an error is marked with "E" like
. Check on the power unit that has an
error. First, leaving the energization setting as is, select OK. All the power units are restarted.
3.
When the error message window is displayed again, press the ENTER key.
4.
If the error still exists, press the [PU-ON] soft-key again to open the power unit energization
setting window.
5.
Check on the power unit that has an error. If the same power unit still has an error, it may be
faulty. Move the cursor to this power unit, then press the ENTER key. The check mark on this
power unit is removed, showing
. If a different power unit has an error, leave the check
mark as is. Select [OK].
6.
The power unit energization setting is updated, and only the power units with the check mark
are restarted. When the error message window is displayed again, press the ENTER key.
7.
Repeat the step 4 and later ones until errors are resolved. After the error is resolved, the product
can be continuously used using the power units with the check mark.
Содержание KP3000GS
Страница 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 2: ......
Страница 3: ...DA00033845 005 PROGRAMMABLE AC DC POWER SOURCE KP3000GS INSTRUCTION MANUAL...
Страница 4: ......
Страница 5: ......
Страница 19: ......
Страница 20: ...KP3000GS 1 1 Outline 1 1 Overview 2 1 2 Features 2...
Страница 23: ......
Страница 197: ......
Страница 198: ...KP3000GS 179 5 Description of Screen and Menu 5 1 Screen Configuration 180 5 2 Menu Composition 184...
Страница 207: ......
Страница 208: ...KP3000GS 189 6 Remote Control 6 1 Communication Interface 190 6 2 Remote Local State Switching 199...
Страница 219: ......
Страница 224: ...KP3000GS 205 8 Troubleshooting 8 1 Error Messages and Error Handling 206 8 2 When a Failure is Suspected 213...
Страница 243: ......
Страница 275: ......
Страница 277: ......
Страница 278: ......
Страница 279: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...