116
Approximately 185 hours at a
rate of 2 mL/hr
5 years life
Graphic LCD, 4 text fields, 16 characters
per field
0.1 to 50 mL/hr
Up to 125 mL/hr
± 5% for volume delivery >1 mL under the
following conditions:
• Temperature 55° to 90° F
• Back pressure to 300 mmHg (5.8 PSI)
±10% for volume delivery >1 mL under
the following conditions:
• Temperature 45° to 55°F
• Temperature 90° to 104°F
• Back pressure to 750 mmHg (14.5 PSI)
The maximum volume that may be infused
under Single Fault Conditions is 0.8 mL.
Recommended administration set usage not
to exceed 72* hours.
*follow standards of practice for set changes
Proprietary
Moog Curlin Infusion
non-
DEHP, PVC administration sets available in
different configurations.
Non-filtered microbore sets ± 4 mL.
See individual administration set label for
specific fill volume
Rate
0.1 mL/hr
1 mL/hr
10 mL/hr
50 mL/hr
Time to Alarm
<25 minutes
< 4 minutes
<20 seconds
<10 seconds
ITEM
Operating Time with Full Charge
Batteries
Internal Memory Battery
Display
Flow Rates
Bolus Rates
Volume Accuracy
Administration Sets
Fill Volume for Administration Sets
Up Occlusion Detection Time
Microbore tubing with occlusion below
spike
Power Requirements
2 C-cell single-use batteries
AC Power
SPECIFICATION
Содержание CURLIN INFUSION PainSmart
Страница 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Страница 39: ......
Страница 49: ......
Страница 113: ......
Страница 121: ......
Страница 125: ...118...
Страница 126: ...119 Technical Specifications...
Страница 127: ...120...
Страница 128: ...121 Technical Specifications...
Страница 129: ...122...
Страница 130: ...123 T...
Страница 131: ...124...
Страница 137: ......
Страница 139: ......
Страница 140: ......
Страница 141: ...Rev I 360 9043...