• Refer to the LED indicators on the pump to determine if the battery pack has been fully
charged or is in a low battery condition.
• The battery pack should be unplugged from the pump when recharging. The AC adaptor
cannot simultaneously power the pump and charge the battery pack. If AC operation is
desired, connect the adaptor directly to the pump.
Notice
The Rechargeable Battery Pack DFU is a supplement to the Moog Medical Devices Group Pump
User's Manual. All other applicable WARNINGS, CAUTIONS and NOTICES are found in the
User's Manual and should be referred to when using this DFU.
Environmental
Storage Temperature -20ºC to 65ºC; Atmospheric pressure 500hPa to 1060hPa; Humidity 10%
to 90%
Operational Temperature 5ºC to 40ºC; Atmospheric pressure 650hPa to 1020hPa; Humidity
10% to 90%
Directions for Use
Charging the Battery Pack
The Moog Curlin 360-2022Kit (Globtek GTM41076-0609-1.5) and 47684 and 47685 AC
adapters can be used to recharge the battery. A fully discharged battery will recharge in less than
15 hours.
Note: Do NOT use Moog Curlin 360-2026 (Globtek GTM341-7.5-800), Moog Curlin 340-
2026 (Globtek GTM341-7.5-800), or the Moog Curlin 340-2022 (POWDEC AC Adaptor
WM10-6.9V) AC adaptors to recharge the battery.
• If connected, disconnect the battery pack from the pump.
• Connect the AC adaptor to the power port of the battery pack.
Содержание CURLIN INFUSION PainSmart
Страница 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Страница 39: ......
Страница 49: ......
Страница 113: ......
Страница 121: ......
Страница 125: ...118...
Страница 126: ...119 Technical Specifications...
Страница 127: ...120...
Страница 128: ...121 Technical Specifications...
Страница 129: ...122...
Страница 130: ...123 T...
Страница 131: ...124...
Страница 137: ......
Страница 139: ......
Страница 140: ......
Страница 141: ...Rev I 360 9043...