126
Recommended separation distances between portable and mobile RF communications
equipment and the Painsmart™ IOD, 6000 CMS™ and 6000 CMS™ IOD™
The Painsmart™ IOD, 6000 CMS™ and 6000 CMS™ IOD™ is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Painsmart™ IOD, 6000 CMS™ and 6000 CMS™ IOD™ can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Painsmart™ IOD, 6000 CMS™ and 6000 CMS™ IOD™ as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
0.116
0.369
1.16
3.69
11.7
0.12
0.34
1.7
3.7
11.7
0.23
0.74
2.3
7.4
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in metres (m) can
be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorbtion and reflection from
structures, objects and people.
150 kHz to 80 MHz
80 MHz to 800 MHz
Separation distance according to frequency of transmitter m
800 MHz to 2,5 GHz
d
= [ ] √P
3,5
V
1
d
= [ ] √P
d
= [ ] √P
3,5
E
1
7
E
1
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The PainSmart™ IOD , 6000 CMS™ and 6000 CMS™ IOD™ is intended for use in electromagnetic environment specified below. The
customer or the user of the PainSmart™ IOD, 6000 CMS™ and 6000 CMS™ IOD™should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
2 kV for power supply
lines
±
1 kV for input/output
lines
±
1 kV line(s) to line(s)
±
2 kV line(s) to earth
<5 %
U
T
(>95 % dip in
U
T
)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T
)
for 5 s
<5 %
U
T
(>95 % dip in
U
T
)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T
)
for 5 s
±
1 kV line(s) to line(s)
±
2 kV line(s) to earth
±
2 kV for power supply
lines
±
1 kV for input/output
lines
±
6 kV contact
±
8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the PainSmart™ IOD, 6000 CMS™
and 6000 CMS™ IOD™ requires continued
operation during power mains interruptions, it
is recommended that the PainSmart™ IOD,
6000 CMS™ and 6000 CMS™ IOD™ be
powered from an uninterruptible power supply
or a battery.
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment-
Guidance
Содержание CURLIN INFUSION PainSmart
Страница 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Страница 39: ......
Страница 49: ......
Страница 113: ......
Страница 121: ......
Страница 125: ...118...
Страница 126: ...119 Technical Specifications...
Страница 127: ...120...
Страница 128: ...121 Technical Specifications...
Страница 129: ...122...
Страница 130: ...123 T...
Страница 131: ...124...
Страница 137: ......
Страница 139: ......
Страница 140: ......
Страница 141: ...Rev I 360 9043...