59
Programming a PainSmart
pump with Protocol Library
Safety System (PLSS) Version
2 program
Refer to Protocol Library Safety Software manual, part number 360-7042
YES
key is used to verify or enter
NO
key is used to change, erase or leave screen
Turn the pump on, watch self test and power source bar.
Set tubing must be installed for pump to run.
Press the and keys to scroll through the menu, and press the
YES
key to accept/confirm a
field entry and the
NO
key to reject/change a field entry.
Note: Library therapies may have soft or hard limits set for various values.
For PCA therapies: Hard or soft limit settings may be set for Max Basal Rate, Max Patient Bolus
Amt, and Min Bolus Interval. Hard limits may be set for Max Load Dose.
Note:
A warning screen will display if a value is entered that is out of range of the soft limit
setting(s) for a downloaded therapy. The screen will ask the user to override the soft limit value.
Pressing
YES
will accept the new value, example below:
Basics Pr
og
ramming