82
remainder of the infusion.
Note:
If the screen is not acknowledged with a YES or NO key press, the
Air-In-Line sensitivity will be set to 2 mL when the screen is
cleared (i.e. alarm occurs while screen is displayed).
LCD Display:
***ALERT *** AIR SENSOR HAS BEEN SET TO 2 ML
Alert:
Air-In-Line sensitivity will be set to 2mL
Cause:
NO was pressed at the “AIR SENS OFF. Using In Line Filter?” alert screen.
The Air-In-Line sensitivity will be set to trigger an Air-In-Line alarm if
2mL of air is detected.
Resolution:
If an air eliminating filter is being used, press YES when the “AIR SENS OFF.
Using In Line Filter?” screen is displayed. This will disable air detection as
indicated under “AIR SENS OFF. Using In Line Filter?” alert explanation.
LCD Display:
***ALERT *** AMT NOT 0 OR CURRENT DOSE AMT
Alert:
Unable to change clinician dose while clinician dose in progress
Cause:
A clinician dose is in progress, and a new clinician dose value is entered that is
not 0 or the current dose amount.
Resolution:
If the current clinician dose is to be continued, change the “Clin DOSE”
value
back to the original value and accept the field and menu.
If the current clinician dose is to be terminated, which will allow a new
clinician dose to be programmed, enter a value of 0 in the “Clin DOSE” field
and accept the field and menu. The next time that the Options menu is
accessed a new “Clin DOSE” value can be entered.
LCD Display:
***ALERT *** AMT TBI GREATER THAN BAG VOLUME
Alert:
Value entered is greater than value in “Bag Vol” field.
Cause:
Occurs when the amount the entered for the “Amt TBI” or “Vol TBI” field is
greater than the value in the bag volume field.
Resolution:
Recheck the data entry.
LCD Display:
***ALERT *** AMT TBI GREATER THAN BAG VOLUME. YES to
Continue NO to Return
Alert:
Insufficient volume
Содержание CURLIN INFUSION PainSmart
Страница 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Страница 39: ......
Страница 49: ......
Страница 113: ......
Страница 121: ......
Страница 125: ...118...
Страница 126: ...119 Technical Specifications...
Страница 127: ...120...
Страница 128: ...121 Technical Specifications...
Страница 129: ...122...
Страница 130: ...123 T...
Страница 131: ...124...
Страница 137: ......
Страница 139: ......
Страница 140: ......
Страница 141: ...Rev I 360 9043...