32
• Disconnect the battery from the AC adaptor.
Note: If there is a fault detected while charging the battery, the LED will either flash yellow
or not illuminate. Most faults are due to either the AC adaptor becoming unplugged from
the power outlet or the battery pack being at a location where the temperature is not between
5 and 40 degrees celsius. If the fault condition persists, then the battery pack should be
returned to Moog Medical Devices Group for replacement of the battery.
Using the Battery Pack to Power the Pump
• Make sure fully charged C-cell batteries are loaded in the pump.
• Connect the battery pack to the AC adaptor port on the pump.
• Operate the pump in the desired infusion mode.