84
LCD Display:
***ALERT *** BOLUS REQUESTED
Alert:
Not time for bolus
Cause:
BOLUS button or key is pressed before the programmed time interval
between
bolus doses have elapsed.
Resolution:
Advise patient of time limits between bolus doses per agency protocols.
LCD Display:
***ALERT *** CALCULATED AMT OUT OF RANGE
Alert:
Invalid calculated amount for infusion
Cause:
The calculated amount for the infusion, based on the values entered for rate
and time, exceeds either the pump limit or the value entered for bag volume.
Resolution:
Recheck data entry for Rx.
LCD Display:
***ALERT *** CALCULATED RATE OUT OF RANGE
Alert:
Invalid calculated rate of infusion
Cause:
The calculated rate for the infusion, based on the values entered for Volume/
Amount and Time, exceeds either the minimum or maximum rate limit of the
pump or the maximum rate programmed in medLMTS
™
.
Resolution:
Recheck data entry for Rx.
LCD Display:
***ALERT *** CALCULATED TIME OUT OF RANGE
Alert:
Invalid calculated time
Cause:
The calculated time for the therapy, based on the values entered for Volume/
Amount and Rate, exceeds the therapy time limit of 288:00 HH:MM (167:59
HH:MM Software Versions below R37C).
Resolution:
Recheck data entry. The time calculated by the pump cannot exceed the
pump range.
LCD Display:
***ALERT *** CHANGES REQUIRE REPROGRAMMING. YES to
Continue NO to Return
Alert:
Changes to the data will require re-entry of prescription.
Cause:
Attempted to change the Pre-Prescription Screen prior to start of infusion.
Resolution:
If a change to the programmed prescription is required, press “YES”.
Note:
Selecting YES will clear the entire prescription and require re-entry of
the prescription.
If no change is required, press NO to preserve the current
prescription and return back to the Pre-Prescription screen.
LCD Display:
***ALERT *** DATE/TIME CHANGE NOT ALLOWED
Alert:
Not allowed to change date or time.
Cause:
Attempt made to modify system clock while a therapy/delay is in progress.
Resolution:
If the therapy in progress is to be continued, wait until the infusion is
complete and then change the date and time.
Содержание CURLIN INFUSION PainSmart
Страница 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Страница 39: ......
Страница 49: ......
Страница 113: ......
Страница 121: ......
Страница 125: ...118...
Страница 126: ...119 Technical Specifications...
Страница 127: ...120...
Страница 128: ...121 Technical Specifications...
Страница 129: ...122...
Страница 130: ...123 T...
Страница 131: ...124...
Страница 137: ......
Страница 139: ......
Страница 140: ......
Страница 141: ...Rev I 360 9043...