69
Note:
The amount of fluid used in the priming process will be deducted from the Bag Volume
Total but will NOT be counted in the amount infused because this amount is not to be
delivered to the patient. The bag volume should still be equal to or greater than the Amount
TBI before you can proceed with the infusion. If it is not, an alert will appear stating, “AMT
TBI GREATER THAN BAG VOL.” You have the option to continue by pressing the YES/
ENTER key or return to the Run Options screen by pressing the NO key.
Options
Once the prescription is programmed, there are more customizing features available for each therapy.
Press the OPTIONS key from the Run Options screen to access these options. The OPTIONS
menu can also be accessed while the pump is running by pressing the Help/Options key.
Note.
After the infusion has started, any changes made through the OPTIONS menu involving:
• “Clin DOSE”,
• “SHFT TOT”
• “HRLY TOT”, and
• “CLR HRLY”
will return you directly to the “ACCEPT OPT?” field.
The following screen displays the PCA Therapy Options available. When you open this menu, use
the up and down arrows to move the highlight bar to the desired option. Lock settings may require
an access code before the option can be changed. Any new setting, except “CLEAR HOURLY”
screen acceptance and “SHIFT TOTAL” clear screen acceptance becomes effective when the
Options screen is exited by pressing YES/ENTER at the ACCEPT OPT? field.
P C A O p t i o n s M e n u
Note:
The pump’s default settings are indicated above in the
left
screen.
1. LOCK: When the lock is set to “OFF,” the clinician has access to all of the pump functions.
After the pump is completely programmed, change the lock to meet the security needs of this
therapy. There are four choices: “OFF”, “1”, “2”, or “3”. Each level higher limits the available
options. To change the lock setting, press the NO/CHANGE key until the correct lock setting
appears. Then, press YES/ENTER key to accept the new lock selection. Once the pump has been
set into a locked condition, any subsequent request to change the lock will
PCA T
her
ap
y
Содержание CURLIN INFUSION PainSmart
Страница 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Страница 39: ......
Страница 49: ......
Страница 113: ......
Страница 121: ......
Страница 125: ...118...
Страница 126: ...119 Technical Specifications...
Страница 127: ...120...
Страница 128: ...121 Technical Specifications...
Страница 129: ...122...
Страница 130: ...123 T...
Страница 131: ...124...
Страница 137: ......
Страница 139: ......
Страница 140: ......
Страница 141: ...Rev I 360 9043...