• Beaker with clean, fresh water or IV container.
Note:
If the administration set contains a filter, sterile saline or sterile water should be used.
Maintenance between Patients
1. CLEAN AND INSPECT PUMP FOR DAMAGE:
a. Clean the pump with a soft cloth dampened with a designated antiseptic, disinfectant
solution or a solution of household bleach diluted 9:1 with water. Be sure to clean all
surfaces, including the pumping chamber.
b. Inspect pump for damage. Check all case surfaces for cracks. Open the pumping chamber
door. Examine the finger seal to confirm that no cuts or tears exist. Confirm that the door
is well aligned by closing it with no unreasonable interference. Verify that the power and
bolus/data connectors are securely mounted on the case (no loose nuts). Verify that all
labels are legible and in good condition. Check that keypad is securely attached to case and
that no corners are pealing up. Verify that the battery door is in place and that the battery
contacts in the battery compartment are not deformed.
Refer to the picture below when executing the following instructions.
2.
Volumetric Delivery Test:
a. Install a new primed Administration Set in the pump and latch the door.
b. Connect the up-stream side of administration set to a water source 12 /-2
Cleaning, Maintenance
Содержание CURLIN INFUSION PainSmart
Страница 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Страница 39: ......
Страница 49: ......
Страница 113: ......
Страница 121: ......
Страница 125: ...118...
Страница 126: ...119 Technical Specifications...
Страница 127: ...120...
Страница 128: ...121 Technical Specifications...
Страница 129: ...122...
Страница 130: ...123 T...
Страница 131: ...124...
Страница 137: ......
Страница 139: ......
Страница 140: ......
Страница 141: ...Rev I 360 9043...