Cause:
The Library Protocol units are not valid for the pump configuration.
Resolution:
Select a different Library Protocol or revise the Library Protocol.
LCD Display:
***ALARM *** INVALID ROUTE FOR PUMP CONFIGURATION
Alarm:
Invalid Route for Pump Configuration
Cause:
The Library Protocol Route is not valid for the pump configuration
Resolution:
Select a different Library Protocol or revise the Library Protocol.