67
This field will not be accepted with the YES/ENTER key until the rate is adjusted to fall
within pump’s range. When the rate is correct, use the YES/ENTER key to accept, and
move to the next field.
3. Pt. BOLUS (Patient Bolus Dose): Use the numeric keys to enter the prescribed bolus (patient
demand dose) and when correct, press the YES/ENTER key to accept and move on to the next
field. (see Figure 5.1.)
Note:
If a bolus dose entered falls out of the range of the pump, the above “Greater than Max
Field Limit” or “Less than Min Field Limit” alert will also appear and display the limit.
4. BOLS INT (Patient Bolus Interval): This field allows the clinician to enter the minimum
amount of time (from 0 to 60 min) prescribed between patient-activated bolus dosing. Enter the
time using the numeric keys and, when correct, press the YES/ENTER key to accept and move
the cursor to the next field.
5. # BOLS/hr (Number of Patient Bolus Doses per hour): Using the numeric keys, enter the
number (0 to 15) of boluses allowed per hour and, when correct, press the YES/ENTER key to
accept and move to the last field.
6. DONE?: If the programming is not done, use the up or down arrow keys to locate the field
to be changed, use the numeric keys to enter the corrected information and press YES/ENTER
to accept the new data. If all the fields are acceptable, at the DONE? prompt, press the YES/
ENTER key to move on to the next screen.
Before the infusion is started, it is important to consider a few
remaining issues, such as checking the administration set to be sure it
was primed or purged of all air, or perhaps the pump may need to be
put into a security lock setting for additional safety.
The following screen offers you four choices:
R u n O p t i o n s S c r e e n
Note that you do not need to answer yes or no questions or enter data as previously requested. This
screen requires only that you press the corresponding key on the keypad to begin the requested
function. The following section explains each selection beginning with Prime:
Prime
The IV bag or syringes and administration set should be inspected and be free of all air and air
bubbles before they are connected to the access site. Priming may have been done by gravity when
the administration set was first connected to the bag (see Chapter 2). However, if it was not done
then or if there are any remaining air bubbles to be removed, the pump’s “PRIME” function will
assist in this process. To prime the set by using the pump, follow the directions below.
Note:
The lock setting may restrict the ability to prime using the pump.
IMPORTANT
INFORMATION
PCA T
her
ap
y
Содержание CURLIN INFUSION PainSmart
Страница 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Страница 39: ......
Страница 49: ......
Страница 113: ......
Страница 121: ......
Страница 125: ...118...
Страница 126: ...119 Technical Specifications...
Страница 127: ...120...
Страница 128: ...121 Technical Specifications...
Страница 129: ...122...
Страница 130: ...123 T...
Страница 131: ...124...
Страница 137: ......
Страница 139: ......
Страница 140: ......
Страница 141: ...Rev I 360 9043...