57
Resuming a Therapy
1. RESUME—If a therapy is interrupted and the pump is turned off or paused before completion
of the infusion, you may select “RESUME” by moving the cursor to this field and pressing
the YES/ENTER key. A notification screen displays, and a prompt is given to run the pump.
At this time, you can continue to infuse the prescription from the exact point at which it was
interrupted.
Note:
Patients should be taught proper time frames or limitations for leaving their IV bags or
medication reservoirs hanging during an interrupted state and, if indicated, to change
to a new bag according to the healthcare provider’s policies and drug manufacturer’s
instructions.
Repeating a Pre-Programmed Therapy
2. REPEAT Rx—Use this feature if an infusion is complete and you wish to repeat the same
prescription over again or if a therapy is programmed and the pump is turned off before the
infusion is started. Select this field by moving the cursor to REPEAT Rx, and press the YES/
ENTER key.
If the pump has been programmed for more than one therapy and you wish to select a therapy
at this time, move the cursor to “REPEAT Rx,” and press the YES/ENTER key. This action will
take you to the Select Therapy Screen and allow you to choose which of the programs you now
wish to repeat.
Note:
If you are choosing a therapy that is different than the last one selected, Shift Totals and
Hourly Totals will be cleared.
If only one therapy is programmed and “REPEAT Rx” is selected, the cursor will automatically
advance through the prescription menus. You can stop the auto-advance function by pressing
any key (except ON/OFF) and then complete the confirmation of the prescription by pressing
the YES key at the remaining fields.
Note:
REPEAT Rx
does not allow any changes to the Pre-Prescription Menu.
If changes need
to be made in Pre-Prescription selections (units, concentration, etc.) select “New Program”
and reenter the entire prescription.
The lock settings affect the use of “REPEAT Rx.” Lock 1 allows REPEAT with changes to the
prescription screen only. Lock 2 allows REPEAT but does NOT allow the user to make changes
in the prescription. Lock 3 does not allow REPEAT. (See lock setting table.)
If a therapy has not completed and the “REPEAT Rx” function is selected, the following screen
will display:
Basics Pr
og
ramming
Содержание CURLIN INFUSION PainSmart
Страница 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Страница 39: ......
Страница 49: ......
Страница 113: ......
Страница 121: ......
Страница 125: ...118...
Страница 126: ...119 Technical Specifications...
Страница 127: ...120...
Страница 128: ...121 Technical Specifications...
Страница 129: ...122...
Страница 130: ...123 T...
Страница 131: ...124...
Страница 137: ......
Страница 139: ......
Страница 140: ......
Страница 141: ...Rev I 360 9043...