114
c.
Up Stream Occlusion
- Resume the infusion pumping then pinch off or close the up-stream
side of the administration set. Confirm that the pump goes into an Up Occlusion alarm.
Open or clear the up stream occlusion, press PAUSE then RESUME the infusion.
d.
Door Sensor
- With infusion in progress open the door latch and observe “ALARM DOOR
OPEN OR SET NOT PROPERLY INSTALLED.” Press PAUSE and then “YES to
RESUME” on the Pause Menu.
e.
Set Sensor
- Remove the administration set from the pump and reinstall it so that the flow
stop and tubing guide are not in the pumping chamber. Close the door latch and start an
infusion by pressing the RUN key. Observe “***ALERT *** DOOR OPEN or SET NOT
PROPERLY INSTALLED” will occur.
f.
Air In Line Sensor
- Reinstall the administration set properly and resume the infusion.
From the Options menu set the AIL SENS (AIR IN LINE sensitivity) to 0.1mL. With the
pump running introduce an Air Bolus of approximately 0.5mL into the fluid path at the
input side of the pump. Once approximately 0.1mL of air has past the AIL detector, an
“ALARM AIR IN LINE” will occur. Pause and prime the air bolus past the AIL sensor, then
resume the infusion.
g.
Auxiliary Alarm
- With the pump running observe the Vol INF (volume infused on the run
screen) and disconnect the pump from all power sources (unplug the AC Adaptor
eliminator and remove one C-cell battery.) Observe that the Auxiliary Alarm System is
activated, with the red LED and beeper on constant (not intermittently beeping). With
the battery still out of the pump, press the ON/OFF button to turn off the auxiliary
alarm. Replace the battery in the battery compartment and turn on the pump. Resume the
therapy and confirm that the infusion continues where it stopped due to Auxiliary Alarm
occurrence.
4 SET BioMed CLINICIAN MENU ITEMS:
a. Turn the pump off and back on. While pump is performing self tests confirm that the
pump is set to the correct date and time. If date and time are not correct, program the
correct date and time from the BioMed/Clinician menu.
Note:
Setting of the date and time is not allowed if “RESUME” (infusion is still active) is
available on the Pause Menu. Use the “CLEAR Pt Data/Rx” option on the BioMed/
Clinician menu to clear the prescriptions and end the infusion. The date and time can
then be programmed.
b. From the Program/BioMed Setup Menu select “BioMed Setup.”
c. At the Enter Code Menu, enter the Clinician Code.
d. (Optional) Erase the patient history file using the Clear Hx Log Menu.
e. Erase any previous Programmer ID, Patient ID, Rx Number, Invoice Number, Lock Level
setting, and prescriptions using the Clear Pt Data/Rx Menu.
f. Observe the Pump Maintenance Date, top of the BioMed/Clinician Menu, to insure
adequate time for the next use by patient.
g. Sign, date and return the check list to your record file. Pumps that pass can be used for
new patients. Pumps that fail are to be returned for service.
Cleaning, Maintenance
Содержание CURLIN INFUSION PainSmart
Страница 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Страница 39: ......
Страница 49: ......
Страница 113: ......
Страница 121: ......
Страница 125: ...118...
Страница 126: ...119 Technical Specifications...
Страница 127: ...120...
Страница 128: ...121 Technical Specifications...
Страница 129: ...122...
Страница 130: ...123 T...
Страница 131: ...124...
Страница 137: ......
Страница 139: ......
Страница 140: ......
Страница 141: ...Rev I 360 9043...