87
2
Is the
monitor display outside
the range 4 kHz to
16 kHz?
Recalibrate in accordance with
the procedure in [4. Calibration].
Is the situation
unchanged?
Repair or replace the torque
transducer.
Yes
Yes
Contact us and describe the issue
and situation in as much detail as
possible.
Start
Is F-54 set to 0?
Are the
frequency output
specifications of the torque
transducer connected
10 kHz ±5 kHz?
Yes
Yes
Set F-54 to 1.
The frequency
output specifications of the
torque transducer connected are
240 kHz ±120 kHz.
Set F-54 to 0.
No
No
No
No
Is the
input cable mixed in with
other power
cables?
Is the
stator mounting stand
grounding faulty?
Wire the input cable separately.
No
Reinforce the stand grounding.
No
Contact us and describe the issue and
situation in as much detail as possible.
3
Yes
Yes
Содержание OPT-564
Страница 1: ...DIGITAL TRANSMITTER OPT 564 Instruction Manual EN294 1924B ...
Страница 2: ......
Страница 12: ......
Страница 80: ...68 Frequency Output Shield Frequency counter or other device ...