24
Zero/Span Fine Adjustment Function
This function fine-adjusts the zero point and span point if there are discrepancies between the actual
torque and the reading displayed.
Selecting Zero Point Fine-adjustment Mode
Zero point fine-adjustment mode can be selected from measurement mode as follows.
appears in the upper part when the
key is
depressed for approximately 2 seconds in measurement mode.
The upper display reads
when the
key is
pressed twice.
The upper display reads
and the reading on the
lower display flashes, indicating zero point fine-adjustment mode has
been selected when the
key is pressed.
The lower display displays the torque value currently entered.
Set the torque transducer to the initial torque state, and adjust the
display to "zero".
: Increments the torque.
Hold down the key to increment the torque continuously.
: Decrements the torque.
Hold down the key to decrement the torque continuously.
: Cancels the setting and returns to measurement mode.
: Registers the value displayed and proceeds to the next
step.
Press the
key after setting.
is displayed after zero point fine-adjustment.
Press the
key to exit zero point fine-adjustment mode.
The system switches to measurement mode and the data set is
registered.
Measurement
mode
Current torque
value
(twice)
Measurement
Mode
Содержание OPT-564
Страница 1: ...DIGITAL TRANSMITTER OPT 564 Instruction Manual EN294 1924B ...
Страница 2: ......
Страница 12: ......
Страница 80: ...68 Frequency Output Shield Frequency counter or other device ...
Страница 114: ...102 External View The individual component dimensions are as follows Front Rear Units mm Panel cut out dimensions Side ...