43
CHECK Value
Sets the analog output when the
key is held down. (F-20)
Setting range: 0 to 20
Output:
(Maximum analog output – Minimum analog output)/20 × Preset value
For example, if the maximum analog output is 10 V and the minimum analog output is 0 V, the analog
output when the
key is pressed will be (10 V – 0 V) / 20 × 20 = 10 V.
By default, this is set to [16].
Rotation Detection Input Signal Format
The rotation speed can be detected by connecting a rotation detector (MP-9820).
The rotation speed detection method used for the rotation detector can also be set. (F-24)
Setting range: 0 to 3
0:
Off
Rotation detection is not performed.
1:
Measure using 2 rotation detectors
Rotation is detected using the rotation detector 2 input signal. The rotation
direction is detected using the signal phase.
2:
Measure using 1 rotation detector. The rotation direction is detected using the
external input signal.
Rotation is detected using the rotation detector 1 input signal. The rotation
direction is detected using the external input signal (Rotation POL.).
3:
Measure using 1 rotation detector. Rotation direction cannot be detected.
Rotation is detected using the rotation detector 1 input. The rotation direction
cannot be detected.
By default, this is set to [0].
No. of Rotation Detection Gear Teeth
The number of gear teeth for rotation detection can be selected. (F-25)
Setting range: 0 to 2
0:
120 teeth
1:
240 teeth
2:
360 teeth
By default, this is set to [0].
When using this, be sure to set this to match the number of teeth on the gear wheel used. The standard
rotation detection gear wheel on our torque meter has 120 teeth.