20
Zero Calibration
Register the input at the zero point.
Select the zero calibration method.
To calibrate using the torque value (initial torque state), press the
key. Use the method in a).
The torque transducer is zero calibrated at the initial torque state.
To calibrate by entering the torque value for the torque transducer, press the
key, then press the
key.
Use the method in b). Calibrate zero by entering the zero point torque (frequency) for the torque
transducer.
a)
Calibrating using the torque value (initial
torque state)
Set the torque transducer to the initial torque state.
When
is displayed on the lower
display, press the
key. The zero point is
registered and
is displayed.
b)
Calibrating by entering the torque value for the
torque transducer
Press the
key to display
in
the lower display, and then press the
key to
display the value.
The value displayed will be the output frequency for
the torque transducer at the initial torque state
previously registered.
Set the torque value (frequency) corresponding to
the zero point in 1 Hz intervals. (When F-54 is set to
1, the units are 10 Hz.)
/
: Alter the value of the digit to be altered.
/
: Select the digit to be altered.
: Cancels the setting and returns to
measurement mode.
: Registers the value displayed and
proceeds to the next step.
Press the
key after setting.
With no
torque on
the torque
transducer,
press the
key.
To span calibration
Zero
calibration
in progress
Hz
Registered torque transducer
output value
After
is
displayed,
press the
key.
Hz
Set torque transducer output value