72
RS-232C Communication Setting Items
Operating mode (F-30)
The RS-232C communication operation mode can be selected as [Command mode] or
[Stream mode].
By default, [Command mode] is selected.
Stream output target (F-31)
The data to be output in stream mode can be selected as [Torque display], [A/D data], or [Rotation
speed display]. By default, [Torque display] is selected.
Baud rate (F-32)
The baud rate can be set as [1 200 bps], [2 400 bps], [4 800 bps], [9 600 bps], [19 200 bps],
[38 400 bps], [57 600 bps], or [115 200 bps].
By default, [9 600 bps] is selected.
Data bit length (F-33)
The output data bit length can be selected as [7 bit] or [8 bit].
By default, [7 bit] is selected.
Parity (F-34)
The parity bit setting can be selected as [No], [Odd], or [Even].
By default, [Odd] is selected.
Stop bit (F-35)
The stop bit setting can be selected as [1 bit] or [2 bit].
By default, [1 bit] is selected.
Terminator (F-36)
The terminator setting can be selected as [CR] or [CR+LF].
By default, [CR+LF] is selected.
Decimal point (F-37)
The decimal point for output data can be selected as [No] or [Yes].
By default, [No] is selected.
Communication is performed in measurement mode only. An error command will be sent in all other
modes.
The OPT-564 does not use flow control.
The CTS/RTS signal is not used.
X flow control is not used.
Communication is interactive.
The power must be turned off and then back on again in order to refresh changes to settings.