17
Calibration Procedure
Calibration in Measurement Mode
Step 1
Connect the torque transducer to the equipment.
Step 2
Turn on and wait for approximately 10 minutes for the torque
transducer and equipment to stabilize.
Step 3
Select calibration mode.
Step 4
Select the minimum torque units.
Step 5
Set the maximum torque value that can be measured by the
torque transducer.
Step 6
Set the actual torque value applied to the torque transducer.
Step 7
Calibrate the zero point with the torque transducer at the
initial torque state or by setting the zero point torque for the
torque transducer.
Step 8
Calibrate the span point with the span torque acting on the
torque transducer or by setting the span point torque for the
torque transducer.
Step 9
Calibration is complete. Perform steps 10 onward as
described below where necessary.
Steps 4 to 9 must be carried out in sequence. The calibration results will not be updated if the
procedure is not completed up to step 9.
The A/Z function is canceled once calibration is completed (up to step 9).
The A/Z function is canceled if step 10, 11, or 12 is performed.
Section 4-4-3
Minimum Scale Division
Setting
Section 4-4-4
Torque Setting
Torque transducer
connection
Step 11
Step 12
Step 13
Step 14
Section 4-4-5
Actual Torque Setting
Section 4-4-7
Span Calibration
Section 4-4-6
Zero Calibration
Step 15
After turning on power
Section 4-4-2
Selecting Calibration
Mode
Ending Calibration
Section 4-9
Decimal Place
Setting
Section 4-10
Units Setting
Section 4-6
Symmetry
Correction
Function
Section 4-7
Linearization
Correction
Function
Section 4-8
Polarity
Change
Function
Step 10
Section 4-5
Zero/Span
Adjustment
Function