Torque analog output check
Is the voltage
output load resistance at least
2 kΩ and the current output load
resistance not more than
510 Ω?
Are the connections
correctly configured?
Check the torque analog output
settings.
Are they set
correctly?
Yes
Yes
Yes
Contact us and describe the issue and
situation in as much detail as possible.
No
No
Connect correctly in accordance
with [2-5. Torque Analog Output
Connection].
Use a load resistance to suit the
specifications.
Set correctly in accordance with
[10. Analog Output].
No