29
The change polarity setting will be temporarily canceled when using the linearization correction
function. (Restored after adjustment)
Linearization correction should be set from the point closest to the zero point.
Linearization correction in the clockwise direction should be performed after symmetry correction.
(The clockwise direction linearization correction setting will be reset when symmetry correction is
performed.)
The A/Z function is canceled once the linearization correction function ends.
The linearization correction function is saved respectively for calibration data 1 to 4.
Polarity Change Function
This is a function for changing the polarity of the torque transducer output.
The default setting is for positive output when torque acts on the transducer in the counterclockwise
direction and negative output when torque acts in the clockwise direction.
The function mode F-19 setting can be altered to reverse the polarity to negative output for torque acting
on the transducer in the counterclockwise direction and positive output for torque acting in the clockwise
direction.
For details of the operating procedure, see "9. Function Mode".
The change polarity setting will be temporarily canceled when calibrating, when using the zero/span
fine adjustment function, when using the symmetry correction function, and when using the
linearization correction function. (Restored after adjustment)
Decimal Place Setting
The decimal point display position can be set.
The decimal point display position can be selected from [None], [1.0000], [10.000], [100.00], and
[1000.0].
By default, [None] is selected. For details of the operating procedure, see "9. Function Mode".
Units Setting
The torque units can be set.
The units can be selected from [N
m], [k N
m], and [V].
By default, [N
m] is selected. For details of the operating procedure, see "9. Function Mode".