Power and Ground Connections
Connect the power supply and ground to the power supply connector board on the rear panel of the
equipment as shown in the following figure.
Use Type D single grounding.
CAUTION
For power and grounding, connect securely, as shown. Use within the rated power range.
Since OPT-564 does not have a power switch, install a circuit breaker, etc. if necessary.
Use Type D single grounding for the equipment. If not, the equipment may be susceptible to noise from
other devices, resulting in malfunctions.
100 V to 240 V AC
(Permissible range:
85 V to 264 V AC)
Type D single
grounding