83
Stream mode
Data is entered as right-aligned data.
The sign of the data is indicated by [-] if negative and [+] if positive.
Data is subject to zero suppression.
If decimal point is enabled, a decimal point is added at the specified location when decimal points
are specified on the calibration screen.
If an overload occurs, [OL] is output to the data unit for torque and [OS] is output for rotation speed.
All blank areas are spaces.
The unit part uses the value set by function number F-02. [0] →N·m, [1] →kN·m, [2] → V
The unit part for rotation speed display is fixed at [0], and the unit part for input torque AD data is
fixed at [Space].
Return (OPT-564 → Host)
2
4
6
8
10
12
14
@
0
0
9
9
Unit
Terminator
2
4
6
8
10
12
14
@
0
0
9
9
-
1
0
Terminator
2
4
6
8
10
12
14
@
0
0
9
9
+
0
.
1
0
Terminator
2
4
6
8
10
12
14
@
0
0
9
9
+
O
L
0
Terminator
2
4
6
8
10
12
14
@
0
0
9
9
-
O
L
0
Terminator
Address
00 to 31
Command No.
Address
00 to 31
Data sign
Data
Address
00 to 31
Address
00 to 31
Address
00 to 31