- 5-27 -
17. After verifying the level, press the [EN-
TER] key.
18. Press the [ ] to terminate the illumination
check. Press the [ ] key and then finish
the level check.
When check the illumination again, press the [ ]
and do the step E of the setting the UV
illuminometer.
•
Make sure to check the level again
after the lamp cools down. An accu-
rate result can not be obtained by UV
lamp heat.
19. Remove the UV illuminometer and BKT1.
• Do not touch the BKT1 and sensor
parts of the UV illuminometer on or
right after checking to avoid from burn
injury. Make sure to remove them after
it cools down completely.
20. Attach the Flashing tray and then press
the [ENTER] key.
21. Close the left maintenance cover.
Detect the origin and back to LOCAL.
UV LEVEL CHECK
CHECK<
>FINISH
UV LEVEL CHECK
SET A TRAY
:ent
PLEASE CLOSE
THE LEFT COVER
PLEASE CLOSE
THE LEFT COVER
UV LEVEL CHECK
CHECK START
:ent
<< LOCAL >>
UV LEVEL CHECK
CHK THE TOOL
:ent