15.5 Current Injection.
If an alarm occurs that blocks further operation of the system, the ‘Info’ dialog will be displayed. For a description of
the occurred alarm
Consult chapter 15.4 Message codes. on page 160.
Operation of the system may be blocked due to the following:
Emergency Stop Button is active.
The emergency stop button is pressed down. Release the button to continue operation of the system.
For more information, please consult the chapter ‘System Assembly’ - section ‘Control Unit’ - ‘Emergency
Stop’.
DC Current Generator temperature.
DC Current Generator temperature monitoring is installed and has detected a temperature exceeding 55 °C.
Please wait for the DC Current Generator to cool down.
Battery voltage.
If the battery voltage is below 12 VDC, please wait for the battery to recharge.
PLC error.
PLC errors regarding the I/O system will prevent further operation. A system restart may be required.
Please contact Megger technical assistance if the problem persists
Ultracapacitor temperature.
An ultracapacitor’s temperature has exceeded 60 °C. Please wait for the ultracapacitor to cool down.
Disk space critically low.
The used disk space has exceeded 98%. Please free up disk space by deleting stored report files.
No DC Current Generator connected.
The system has detected no connected DC Current Generator. Connect a DC Current Generator to resume
operation.