Maintenance
Replace mains fuse
48
GA 5752 1555 DE 28
7
CAUTION!
Property damage!
You may only use fuses of the following type:
• 2 x T 1.6 A H / 250 V for nominal voltage 230 V AC (REF 5752 1554)
• 2 x T 2.5 A H / 250 V for nominal voltage 127 V AC (REF 5752 1559)
11
11
Fig. 32: Replacing fuses
Disconnect equipment plug.
Unscrew fuse carriers (1) using a
screwdriver or a coin.
22
11
11
22
Fig. 33: Replacing fuses
Remove the fuse carriers (1).
Remove the fuses (2) from the fuse
carriers.
Insert new fuses.
Insert fuse holders and screw closed.
Содержание WISTA SP 1070
Страница 1: ...OPERATING INSTRUCTIONS SURGICAL ASPIRATOR MEDAP TWISTA SP 1070 GA 5752 1555 DE 28...