Introduction
Basic requirements
GA 5752 1555 DE 28
13
1
TWISTA SP 1070 127 V; 60 Hz (REF 5752 1559)
• Basic equipment
• Mains cable
• Foot switch
• Filter papers (10 pieces)
• Allen key (REF 5750 4687)
1.4.5
Interface description
1.4.5.1
Hydrophobic bacterial and viral filter
NOTE
The use of a hydrophobic bacterial and viral filter is not necessary if a suitable
hydrophobic bacterial and viral filter is integrated for a specific purpose in the septic
fluid jar of a disposal suction liner.
The hydrophobic bacterial and viral filter protects against contaminants which could be present in
the form of particles or aerosols in the gas drawn in. Moreover, the hydrophobic filter serves as
protection against oversuction; the filter closes off the flow of gas to the product in the event of
oversuction. In its function as bacterial and viral filter, it protects the inside of the aspirator from
the ingress of bacteria and viruses. This ATMOS product (REF 5752 1783) is a hydrophobic
bacterial and viral filter with a pore size of 0.2 μm.
Prerequisites
•
Pore size ≤ 1.0 μm.
• Use a bacterial and viral filters suitable for the particular application.
• The tube connector must match the tube being used.
• Hydrophobic filter must close tightly against water passage at an absolute pressure of up to
10 kPa.
• Observe the direction of flow, if applicable (see note on hydrophobic filter).
1.4.5.2
Vacuum connection tube
The vacuum connection tube is used to connect the aspirator (tube connector or overflow
protection device) and the septic fluid jar.
Technical specifications
• Shore hardness of 60
• Internal diameter of 6-8 mm
• Tube length maximum of 60 cm ±10 cm
• Vacuum resistant down to -95 kPa (may not collapse).
Prerequisites
• The internal diameter of the vacuum connection tube should match the outer diameter of the
tube connector on the septic fluid jar cap of the pump.
The vacuum connection tube will be referred to only as ‘connection tube’ below.
1.4.5.3
Septic fluid jar including septic fluid jar cap
The septic fluid jar is used to collect the septic fluids extracted.
Septic fluid jars with mounted septic fluid jar cap will be referred to as suction set in the following.