![MEDAP WISTA SP 1070 Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/medap/wista-sp-1070/wista-sp-1070_operating-instructions-manual_1759337043.webp)
Cleaning and disinfection
Disinfection
GA 5752 1555 DE 28
43
6
6.2.2
Suitable disinfectants
Only surface disinfectants based on the following combinations of active ingredients may be used
for disinfection:
• Aldehydes
• Quaternary compounds
• Guanidine derivatives
Ingredient group
Active ingredients
Aldehydes
2-ethyl-1-hexanal, formaldehyde, glutardialdehyde, glyoxal,
o-phthaldialdehyde, succinaldehyde
Quaternary compounds
Alkyl-didecyl-polyoxethyl ammonium propionate, alkyl-dimethyl-
alkylbenzyl ammonium chloride, alkyl-dimethyl-ethyl ammonium
chloride, alkyl-dimethyl-ethylbenzyl ammonium chloride,
benzalkonium propionate, benzalkonium chloride (alkyl-dimethyl-
benzyl ammonium chloride, coco-dimethyl-benzyl ammonium
chloride, lauryl-dimethylbenzyl ammonium chloride, myristyl-
dimethyl-benzyl ammonium chloride), benzethonium chloride,
benzyl-dihydroxyethyl-coco-alkyl ammonium chloride, dialkyl-
dimethyl ammonium chloride (didecyldimethyl ammonium chloride),
didecyl-methyl-oxyethyl ammonium propionate, mecetronium-ethyl
sulfate, methyl-benzethonium chloride, n-octyl-dimethyl-benzyl
ammonium chloride
Guanidine derivatives
Alkyl-biguanide, chlorhexidine-digluconate, cocospropylene-
diamine guanidinium diacetate, oliogomeric biguanide,
polyhexamethylene biguanide hydrochloride (oligo-diimino imiodo-
carbonyl imino-hexamethylene, polyhexanide)
Tab. 4:
Active ingredients of disinfectants
6.2.3
Disinfection procedure
Wipe disinfect the product in accordance with the instructions of the disinfectant manufacturer
after every cleaning process.
Ensure that the product is free of disinfectant residue.
Perform visual and functional inspections.
6.2.4
Disinfection procedures
Different disinfection procedures may be used for the various components, depending on the
properties of the materials.
Before disinfection remove contaminations and residues from the parts and dry well.
Components
In solution
1
Wipe / spray disinfection
2
Aspirator basic equipment
X
Foot switch
Mains cable
Cap of bacterial filter
X
X
Float / float cage
Rinsing fluid jar
Connection tube / Suction tube