Safety notes
Product safety notes
18
GA 5752 1555 DE 28
2
DANGER!
Infection hazard!
The bacterial filter paper provides additional protection against contamination of the
ambient air.
Do not operate the aspirator without a bacterial filter paper.
CAUTION!
Property damage due to oversuction!
The product may only be operated with the overflow protection in place as
otherwise oversuction could occur. A hydrophobic filter offers an additional
protection against oversuction. It closes off the flow of gas to the product. Particles
in the gaseous phase may clog the hydrophobic filter.
Use a bacterial and virus filter which also protects the inside of the aspiration from
bacterial and viruses.
CAUTION!
Observe ambient conditions!
If the ambient conditions are undercut or exceeded during transportation, storage
or operation, functionality may be affected.
Conduct a functional check and rectify any deficiencies.
CAUTION!
Property damage!
Excessive exposure of plastic housing components to ultraviolet radiation leads to
premature material fatigue, resulting in breakage.
Protect the product against direct sunlight.
CAUTION!
Property damage!
Proper functioning of the mechanical overflow protection is only assured with the
product in upright position.
Place the product in upright position during operation. Operation is only permitted
when the castors are in a locked position.
CAUTION!
Property damage!
The aspirator may not be lifted or carried using the push bar.