
Cleaning and disinfection
Cleaning
40
GA 5752 1555 DE 28
6
6
Cleaning and disinfection
6.1
Cleaning
6.1.1
General
All the components in the aspirator which come into contact with septic fluid must be cleaned and
disinfected after each use.
DANGER!
Danger to life!
Electric shock!
Remove the mains plug from the socket before cleaning / disinfection.
DANGER!
Danger to life!
Electric shock!
Liquid should never be allowed to enter live parts.
DANGER!
Risk due to incorrect use of detergents and disinfectants!
It is strictly advised to observe the manufacturer instructions regarding how to use
the detergents and disinfectants as well as the valid hospital hygiene rules.
DANGER!
Infection hazard!
Product may be contaminated.
Always wear gloves for cleaning and disinfection.
DANGER!
Infection hazard!
Particles of grime may become encapsulated and lead to the product not reaching
the desired germ-reduction after disinfection.
Before disinfection, the product must be cleaned thoroughly of contamination and
encapsulated particles of grime.
CAUTION!
Improper cleaning and disinfection can cause property damage!
Do
not
use the following products for cleaning and disinfection:
• Products containing alcohol (e.g. hand disinfectants)
• Halogenides (e.g. fluorides, chlorides, bromides, iodides)
• Dehalogenating compounds (e.g. fluorine, chlorine, bromine, iodine)
• Products that may scratch the surface (e.g. scouring agents, wire brushes, wire
wool)
• Standard commercial solvents (e.g. benzene, thinner)
• Water containing iron particles
• Cleaning sponges containing iron
• Products containing hydrochloric acid
Use a soft, lint-free cloth or a soft nylon brush to clean the product.
Содержание WISTA SP 1070
Страница 1: ...OPERATING INSTRUCTIONS SURGICAL ASPIRATOR MEDAP TWISTA SP 1070 GA 5752 1555 DE 28...
Страница 59: ...Notes...