Operation
Suction
GA 5752 1555 DE 28
35
4
4.2.1
Switching on the aspirator
11
22
Fig. 25: Switching on the aspirator
Switch on the aspirator (1).
The green power control light (2) is
illuminated.
4.2.2
Setting the vacuum level
NOTE
Refer to the troubleshooting guide if there is insufficient vacuum or no vacuum at
all.
1
2
Fig. 26: Setting the vacuum level
Setting the vacuum level
Kink the suction tube on the patient side
and hold it closed. Set the vacuum with the
regulating knob (1) and check.
Increasing vacuum
Turn the regulator knob (1) to the right.
Read the value on the vacuum gauge (2).
Reducing vacuum
Turn the regulator knob (1) to the left.
Read the value on the vacuum gauge (2).