Initial operation
Mounting the tubes
30
GA 5752 1555 DE 28
3
3.9.5
Tube connection of overflow protection device with septic fluid jar cap (REF 5752 5432)
11
44
33
22
55
Fig. 20: Mounting the tubes
Plug adapter (REF 5752 2295) (1) onto the
connection tube (2).
Insert the adapter with the connection tube
into the septic fluid jar cap (3).
Attach the suction tube (4) to the patient
side tube connector (5) on the septic fluid
jar cap.
3.9.6
Tube connection for vacuum shift (REF 5752 2049)
The mounting of the tubes is described using the septic fluid jar cap (REF 5750 5362) as an
example.
11
22
Fig. 21: Mounting the tubes
Connect the connection tube (1) to the
lower adapter (2) of the vacuum shift.
Attach the other end of the connection tube
to the tube connector (3) on the overflow
protection device.
Overflow protection device and vacuum
shift are connected by the tube.