Operation
Suction
34
GA 5752 1555 DE 28
4
WARNING!
Risk of infection due to using no or a defective hydrophobic bacterial and viral filter.
Secretions enter the aspirator during aspiration.
Stop using the aspirator. Clean and disinfect the aspirator and have it repaired by a
service technician authorised by ATMOS to do so.
WARNING!
Backflow of aspirated secretion!
In the event of oversuction, the aspirated secretion may flow back to the patient if
there is secretion still left in the suction tube.
Before replacing the septic fluid jar in the event of oversuction or switching off the
vacuum, always remove the tube from the patient first.
CAUTION!
Property damage due to oversuction!
The product may only be operated with the overflow protection in place as
otherwise oversuction could occur. A hydrophobic filter offers an additional
protection against oversuction. It closes off the flow of gas to the product. Particles
in the gaseous phase may clog the hydrophobic filter.
Use a bacterial and viral filter which also protects the inside of the aspirator from
bacteria and viruses.
CAUTION!
Property damage due to foaming!
Foam may be created when extracting secretion. Foam is detrimental to the
functioning of the mechanical overflow protection. This gives rise to the risk of
secretions entering and damaging the aspirator.
Always use a hydrophobic filter and, if possible, a commercially available foam
inhibitor.
NOTE
In the event that the hydrophobic bacterial and viral filter has been tripped, switch
off the aspirator. Empty and clean or replace all parts.
NOTE
Monitor the filling level in the septic fluid jar before and after aspiration and, if larger
volumes are being extracted, during aspiration.
When the maximum filling level is reached, switch off the aspirator and empty the
septic fluid jar.
We recommend keeping an additional septic fluid jar in operational position on an
equipment rail as a spare jar to ensure quick switch to an empty jar.
NOTE
Rinse the suction tube briefly with clean water after each extraction cycle.