Operation
Functional test
GA 5752 1555 DE 28
33
4
4
Operation
4.1
Functional test
Prior to using the system, the operator should check that the product is fully functional and in
good condition.
NOTE
Connecting several septic fluid jars in series can cause delayed suction effect and
reduced suction power.
Prior to each use, carry out the following functionality check:
• All components are properly attached.
• The mains cable is undamaged.
• Components made of plastic or rubber (e.g. control panel film, tube, septic fluid jar cap, septic
fluid jar) are in good condition and show no damage due to ageing.
• Bacterial filter paper is in proper condition.
• The overflow protection device and / or hydrophobic bacterial and viral filter are mounted and
functional.
• The overflow protection device and / or hydrophobic bacterial and viral filter have been proper-
ly cleaned and neither residue nor contamination are present.
• Tube connectors and septic fluid jar cap are tightly seated and do not leak.
• No mechanical forces are acting on the tubes.
• Tubes may not be kinked.
•
Maximum vacuum of approximately −90 kPa is reached within about 20 seconds when the
connection tube is held shut.
• The vacuum can be infinitely variably regulated throughout the entire range.
• The septic fluid jar is attached to the aspirator.
• The aspirator has been properly cleaned and neither residue nor contamination are present.
• Damaged parts have been replaced by new parts.
4.2
Suction
DANGER!
Danger to life!
Electric shock!
Check to ensure that the available mains voltage corresponds with the
specifications on the type plate before connecting the mains plug. Product can only
be separated from the power supply by unplugging at the socket.
DANGER!
Infection hazard!
Risk of bacteria and viruses entering the aspirator.
A bacterial and viral filter protects the inside of the aspirator against contamination
by bacteria and viruses.
Use a bacterial and viral filter which also provides protection against oversuction.