Maintenance
Replace mains fuse
GA 5752 1555 DE 28
47
7
No. Malfunction
Cause
Remedy
7
Aspirator working but
vacuum gauge indicates
no vacuum.
The vacuum gauge is
defective.
Have the equipment repaired by
a service technician authorised
by ATMOS.
8
Reduced / no flow rate
Septic fluid jar cap is not in
the correct position.
Position septic fluid jar cap
properly.
Hydrophobic filter is
clogged (vacuum gauge
indicates vacuum).
Replace hydrophobic filter.
Crack in the tube.
Replace tube.
Seal is contaminated.
Replace seal.
Porous seal of septic fluid
jar cap.
Replace seal.
Bent clamp, septic fluid jar
cap does not close.
Replace septic fluid jar cap.
Septic fluid jar is full,
mechanical overflow
protection is closed
(vacuum gauge indicates
vacuum).
Empty the septic fluid jar and
clean or replace septic fluid jar
and mechanical overflow
protection.
Mechanical overflow
protection is contaminated
with exudate.
Clean overflow protection device
or replace septic fluid jar cap.
Tube connection in septic
fluid jar cap is clogged.
Clean the tube connection.
Suction tip is clogged.
Clean suction tip.
The motor is defective.
Have the equipment repaired by
a service technician authorised
by ATMOS.
9
Aspirator has been
exposed to oversuction.
No mechanical overflow
protection device and no
hydrophobic bacterial filter
installed.
Aspirator may no longer be used.
Have the equipment repaired by
a service technician authorised
by ATMOS.
Mechanical overflow
protection device is
sticking; no hydrophobic
bacterial filter used.
Tab. 7:
Malfunctions and troubleshooting
7.5
Replace mains fuse
WARNING!
Electric shock!
Disconnect the electrical plug before changing the fuses.