Initial operation
Overflow protection device/tube connector
GA 5752 1555 DE 28
25
3
3.6.2
Mounting the hydrophobic bacterial and viral filter (REF 5752 1783) in the mechanical
overflow protection device (REF 5752 1775)
The overflow protection device offers the possibility to subsequently connect a hydrophobic
bacterial and viral filter to the mechanical overflow protection. It is required, if the aspirated gas
contains aerosols. It protects the inside of the pump from both humidity and bacteria and viruses.
CAUTION!
Property damage due to foaming!
Foam may be created when extracting secretion. Foam is detrimental to the
functioning of the mechanical overflow protection. This gives rise to the risk of
secretions entering and damaging the aspirator.
Always use a hydrophobic bacterial and viral filter and a commercially available
foam inhibitor.
11
22
Fig. 12: Mounting the hydrophobic bacterial and vi-
ral filter
Inserting the hydrophobic bacterial and
viral filter (with pore size of 0.2 μm) into
overflow protection device
Remove the lid (1) from filter housing by
lifting upwards.
Fit the hydrophobic bacterial and viral filter
(2).
Close filter housing with cover.
3.6.3
Inserting the tube connector
If the equipment is operated with a tube adapter, an overflow protection device must be fitted.
Disposable extraction systems with integrated hydrophobic filter do not require an additional
overflow protection device. The equipment may be used directly with the tube adapter.