Operation
Replacing the bacterial filter paper
36
GA 5752 1555 DE 28
4
4.2.3
Operating the footswitch
The appliance can be transferred to energy-saving standby mode using the foot switch.
11
Fig. 27: Foot switch
Operate the foot switch.
Aspirator is set to stand-by mode.
Yellow light emitting diode (1) is
illuminated.
Operate the foot switch again.
Yellow light emitting diode goes out.
Aspirator is set to operating mode.
4.2.4
Setting the vacuum shift
22
11
Fig. 28: Vacuum shift
The vacuum shift can be set in four
different ways
• Arrow (1) on the rotary switch (2) of the
vacuum shift points to the right: Suction
material will be conducted into the septic
fluid jar on the right.
• Arrow (1) on the rotary switch (2) of the
vacuum shift points to the left: Suction ma-
terial will be conducted into the septic fluid
jar on the left.
• Arrow (1) on the rotary switch (2) of the
vacuum shift points forwards: Suction ma-
terial will be conducted into the septic fluid
jars on the left and right.
• Arrow (1) on the rotary switch (2) of the
vacuum shift points to the back: Vacuum
shift is switched off, aspiration process is
halted.
4.3
Replacing the bacterial filter paper
WARNING!
Infection hazard!
Any and all the parts of the product could be contaminated.
Wear gloves and be absolutely sure to follow the hygiene rules during all cleaning
and reconditioning work.