Taking the unit out of operation
Disassembly
GA 5752 1555 DE 28
39
5
5.3
Disassembly
5.3.1
Detaching tubes
The disassembly of the tubes is described using the septic fluid jar cap (REF 5750 5362) as an
example.
1
2
Fig. 30: Disassembly of tubes
Switch off the aspirator.
Remove the connection tube (1) from the
black connection of the septic fluid jar cap.
Remove the suction tube (2) from the light
chromed tube connector on the patient
side.
5.3.2
Removing the overflow protection device
NOTE
Avoid damages to the edge of the float.
44
33
11
22
55
Fig. 31: Removing the overflow protection device
Remove cap (1) from filter housing
upwards.
Remove the hydrophobic bacterial and viral
filter (2).
Screw off cap (3) from the overflow
protection device.
Detach float cage (4) and remove float (5).