Initial operation
General
GA 5752 1555 DE 28
19
3
3
Initial operation
3.1
General
WARNING!
Risk of injury!
ATMOS products may be used only when fully functional.
Check to ensure that the ATMOS product is fully functional and in good working
order prior to use. ATMOS recommends always having an alternative suction
option ready. This enables aspiration even in the event of product failure.
WARNING!
Infection hazard!
Contaminated components may endanger the health of staff and patients.
Ensure the product is prepared as per hygiene standards before using it for the first
time.
CAUTION!
Property damage due to oversuction!
The product must only be operated with an overflow protection device.
CAUTION!
Property damage due to material failure!
Do not exceed the permissible overall load of 10 kg of each of the equipment rails
of the trolley.
NOTE
Different septic fluid jar holders and septic fluid jar caps can be fitted to the
equipment rails of the trolley. Be absolutely sure to observe the operating
instructions for all the products used in the configuration.
The scope of delivery includes these operating instructions, as well as the individual components
in accordance with the ordered product versions [
Page 12].
Remove the product from its packaging and check the shipment for completeness and to ensure
the scope of delivery is intact.
3.2
Scope of delivery
• Basic equipment,
• Two stands, each with 2 castors (braked),
• Mains cable,
• Foot switch,
• Two tube holders,
Installation material:
• Eight screws,
• Four lockwashers,
• Four plastic discs,
• Four blind plugs,
• Allen key.