Introduction
Basic requirements
12
GA 5752 1555 DE 28
1
Contraindications:
The TWISTA SP 1070 may not be used for the following purposes:
• Outside the medical sector
• In the home care sector
• Being operated directly by the patient
• For vacuum extraction
• For smoke evacuation
• Without smoke evacuation filter, when aggressive vapours are
also generated during the suction of liquids due to easily volatile
components (e.g. when using iodine as disinfectant)
• For the aspiration of flammable or explosive liquids
• In areas prone to explosion risks (AP-M, APG areas)
• For drainages in the low vacuum range (e.g. thoracic and wound
drainage)
• In the standard equipment version in heart surgery and in op-
erations on the central nervous system. This requires separate
equipment providing reliable protection against equipotential
bonding between the patient and contacting, fluid-filled metal
parts.
The product is:
Active
Sterility:
No sterile product
Single-use product /
reprocessing:
The device and parts of the accessories are reusable. For
information on reprocessing, cleaning and disinfection, please see
the operating instructions.
1.4.3.1
For use in cardiac surgery and surgery of the central nervous system (CNS)
The basic version of the product is a BF protection class device. The unit must be configured to
be CF-proof in order to be used for cardiac or CNS applications.
The following guidelines are therefore imperative:
• A hydrophobic bacterial filter must be fitted between the septic fluid jar and the pump inlet
(REF 5752 1783).
• All touchable components of the septic fluid jar must be made of non-conductive materi-
al (plastic), for example, cap plug (REF 5752 0184), Serres
®
disposable suction liner 3 l
(REF 310.0401.0) and Serres
®
outer container 3 l (REF 310.0403.0).
• Only CF-proof suction lances made of a plastic that does not conduct electricity may be used.
• All other equipment parts and accessories must also be CF-proof.
1.4.4
Versions
These operating instructions apply to the versions listed below:
TWISTA SP 1070 230 V; 50 Hz / 60 Hz (REF 5752 1554)
• Basic equipment
• Mains cable
• Foot switch
• Filter papers (10 pieces)
• Allen key (REF 5750 4687)