Maintenance
Malfunctions and troubleshooting
46
GA 5752 1555 DE 28
7
Suggestion:
No. Inspection
Defects are present
No defects
1
Has the product been cleaned
and disinfected according to
the hygiene guideline?
Do not use the product any
longer.
Clean and disinfect the product
according to the guidelines.
Comment:
2
Are there cracks in the
individual components?
Do not use the product any
longer.
Inform service personnel.
Comment:
3
(Space for other tests)
Comment:
Tab. 6:
Visual and functional inspections
7.4
Malfunctions and troubleshooting
No. Malfunction
Cause
Remedy
1
Aspirator does not start
operation, operating
status display is
illuminated.
Vacuum is still present.
Switch off the aspirator, turn
regulating knob to the left, switch
on the aspirator.
The motor is defective.
Have the equipment repaired by
a service technician authorised
by ATMOS.
2
Aspirator does not start
operation, operating
status display is not
illuminated.
Aspirator is in standby
mode.
Switch off standby mode using
the foot switch. The aspirator
starts operating. If the foot switch
is operated again, the aspirator
switches back into standby mode.
3
Aspirator does not start
operation, operating
status display is not
illuminated.
Equipment or mains plug is
not seated properly in the
socket.
Check the equipment and mains
plugs for proper contact.
None or improper voltage.
Check the building fuse, check
the specifications on the type
plate.
The mains fuse is defective. Replace mains fuse.
4
Equipment cannot be
switched on and off.
The electronics are
defective.
Have the equipment repaired by
a service technician authorised
by ATMOS.
5
Aspirator operates but
operating status display is
not illuminated.
LED on the operating
status display is defective.
Have the equipment repaired by
a service technician authorised
by ATMOS.
6
Vacuum cannot be
regulated.
The membrane regulator is
defective.
Have the equipment repaired by
a service technician authorised
by ATMOS.