Cleaning and disinfection
Disinfection
44
GA 5752 1555 DE 28
6
Components
In solution
1
Wipe / spray disinfection
2
Housing of overflow protection
device
X
Hydrophobic bacterial and viral
filter (REF 5752 1783)
3
Disposable
Bacterial filter paper
When in use, must be replaced daily
1
After exposure (as prescribed in the manufacturer's instructions), rinse components thoroughly
with water and dry them afterwards.
2
After exposure (as prescribed in the manufacturer's instructions) remove disinfectant residues
from the components using a moist cloth and dry them afterwards.
3
Replace the hydrophobic bacterial and viral filter immediately if it is discoloured, contaminated
or oversucked. Furthermore, the filter must be exchanged, if the vacuum displayed is above
-0.3 bar / 30 kPa when the vacuum controller is in the ‘max’ position and the suction hose is
open.
Tab. 5:
Disinfection procedures