Maintenance
General
GA 5752 1555 DE 28
45
7
7
Maintenance
7.1
General
Maintenance, repairs and period tests may only be carried out by persons who have the
appropriate technical knowledge and are familiar with the product. To carry out these measures,
the person must have the necessary test devices and original spare parts.
ATMOS recommends: Work should be carried out by an authorised ATMOS service partner. This
ensures that repairs and testing are carried out professionally, original spare parts are used and
warranty claims remain unaffected.
WARNING!
Health hazard!
The aspirator is used in the treatment of patients. The aspirator or parts of the unit
may be contaminated. Prior to returning the aspirator for inspection or repair,
remove the bacterial and viral filter and all tubes and clean and disinfect the
equipment.
7.2
Period tests
At least every 12 months a repeat test of the electrical safety should be performed according to
IEC 62353.
ATMOS recommends an inspection according to the manufacturer‘s specifications.
NOTE
After running for 3000 operating hours a check should be carried out on the unit.
This check must only be performed by authorised personnel.
7.3
Visual and functional inspections
To ensure correct operation, it is necessary to have visual and functional inspections performed
by a trained person prior to each use.
Documentation of the results of the visual and functional inspections is recommended and should
include the date and signature of the person who performed the inspections. The following table
can be used as a template.