Do not cover the ventilator system in any
way, since the functioning of the
equipment may be adversely affected.
•
Do not modify or remove any original
parts.
•
The ventilator system must not be used
during radiotherapy, since this may cause
system malfunction.
•
The ventilator system must not be used
in a hyperbaric chamber.
•
The ventilator system must be kept away
from magnetic resonance imaging (MRI)
equipment.
•
The ventilator system must not be used
with helium or any gas mixture containing
helium.
•
When the ventilator system is used with
MCare Remote Services, use only
network equipment that is safe and
complies with the relevant electrical and
EMC standards such as IEC 60950.
Note:
The network cable is excluded
from this requirement.
•
Only accessories, supplies, and auxiliary
equipment recommended by the
manufacturer should be used with the
ventilator system. Use of any other
accessories, spare parts or auxiliary
equipment may cause degraded system
performance and safety.
•
Use only active humidifiers approved by
the manufacturer of the ventilator system.
Use of non-approved active humidifiers
may result in higher gas temperatures
and increase resistance in filters.
•
During humidification, carefully monitor
the airway pressure. Increased airway
pressure could result from a clogged
filter. Replace the filter if the expiratory
resistance increases or according to the
instructions for the filter, whichever
comes first.
•
Service, repair and installation must only
be performed by personnel authorized
by the manufacturer.
•
The ambient sound needs to be taken
into consideration when setting the alarm
sound level.
•
Always disconnect the patient from the
ventilator system when performing
operations that involve risk for the
patient, such as replacing the O
2
cell,
dismantling etc.
•
Portable RF communications equipment
(including peripherals such as antenna
cables and external antennas) should be
used no closer than 30 cm (12 inches) to
any part of the SERVO-air Ventilator
System including cables specified by the
manufacturer. Otherwise degradation of
this equipment could result.
CAUTIONS:
•
Never leave the patient unattended when
connected to the ventilator system.
•
Before use, make sure the system
version displayed under SYSTEM
STATUS/General corresponds to the
system version described in the User's
Manual.
•
The manufacturer has no responsibility
for the safe operation of the ventilator
system if the requirements specified in
Intended use on page 6 are not
followed.
SERVO-air Ventilator System v2.2, User's Manual
9
|
Introduction
| 1 |
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...