Edit Temporary O
2
increase (%)
To edit the oxygenation concentration:
•
Tap CONFIGURATION/STARTUP
CONFIGURATION/O
2
% and adjust.
The setting entered in the window that opens
determines the default setting for the O
2
BOOST LEVEL and the oxygen level increase
during pre- and post-oxygenation when
DISCONNECTION/SUCTION is used.
It does not affect the O
2
BOOST LEVEL
function in the adult patient category where
the default is 100 %.
Important: The ventilator system must be
restarted to activate the new settings.
Alarms
The alarm settings are the default settings that
the ventilator system is delivered with. These
defaults can be changed.
To access ALARMS:
•
Tap BIOMED/CONFIGURATION/ALARMS.
The following alarm categories can be viewed
and managed depending on the installed
configuration:
1.
Pediatric
2.
Pediatric NIV
3.
Adult
4.
Adult NIV
1
2
3
4
5
6
7
8
9
The following can be adjusted:
7.
Default audio level
8.
Restore default audio level
9.
Restore default alarm limits
Options to restore defaults, cancel or save
changes are available for all alarms.
SERVO-air Ventilator System v2.2, User's Manual
129
|
Service & Settings
| 8 |
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...