6.16 Nebulization
6.16.1 Aerogen nebulizers
The nebulizer is intended for administering
drugs to patients requiring mechanical
ventilation.
The nebulizer operates for a specific period of
time or continuously regardless of ventilation
mode setting. No extra gas volume is added,
i.e. ventilator system settings and values are
not affected.
Refer also to the manufacturer's operating
manual for instructions for use.
6.16.2 Safety guidelines
WARNINGS!
•
The nebulizer must not be left unattended
when connected to a patient.
•
Before administering any medication via
the nebulizer, consult the manufacturer
regarding the appropriateness of
nebulization for that medication. Only use
physician prescribed solutions.
•
Do not use the nebulizer without a filter,
e.g. Servo Duo Guard, connected to the
expiratory inlet of the ventilator system.
Refer to the Servo Duo Guard User's
Manual.
•
During nebulization, carefully monitor the
airway pressure. Increased airway
pressure could result from a clogged
expiratory filter. Replace the filter if the
expiratory resistance increases.
•
The ventilator system accuracy can be
affected by the gas added by use of
other nebulizers than Aerogen nebulizer.
•
During nebulization, check frequently that
aerosol is being generated.
•
Disconnect the SERVO Humidifier/HME
during nebulization; otherwise the
humidifier may become blocked or the
drug may be trapped in the humidifier.
•
To avoid explosion hazards, do not use
flammable agents such as ether and
cyclopropane or aerosolize
alcohol-based medications which can
ignite in oxygen enriched air under high
pressure with this device.
•
To avoid mechanical or electrical
damage, do not drop the nebulizer unit.
CAUTIONS:
•
Before starting the nebulizer, check that
the medication cup is undamaged and
firmly in place.
•
If the nebulizer is used with active
humidification, then the particle size of
the medication may be affected.
•
Perform a function test prior to use to
verify proper operation.
SERVO-air Ventilator System v2.2, User's Manual
103
|
Ventilatory settings and functions
| 6 |
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...