7.2.2 Viewing active alarms
If more than one alarm is active, open the
alarm list in the status bar.
All alarms are displayed by priority in the alarm
list. The list will be continuously updated when
additional alarms occur.
Each alarm is displayed together with a list of
recommended actions.
Tap Alarm history in the alarm list to open the
ALARM HISTORY window.
7.2.3 Responding to alarms
Active alarms can be silenced for two minutes
by tapping Audio pause in the status bar.
Audio paused along with the time remaining
in the silent period are displayed.
2:00
2:00
Audio pause must be tapped for each alarm
that is activated.
If Audio Paused is tapped before the silent
period has expired, then the alarm signal will
be turned on again if the alarm is still active.
Note: The No battery capacity alarm and
high priority technical alarms cannot be
silenced.
To respond to alarms:
•
Tap Audio pause to silence the alarm for
two minutes.
•
Take action to resolve the alarm condition.
Low and medium priority alarms are
automatically reset once the alarm condition
ceases.
High priority alarms are reset automatically or
by confirmation of the alarm.
Refer to section Resetting Check alarms on
114
SERVO-air Ventilator System v2.2, User's Manual
| 7 |
Alarm handling
|
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...