4
5.
To stop nebulization tap
MANEUVERS/NEBULIZATION or the
nebulization shortcut in the status bar.
6.
Tap STOP NEBULIZATION PERIOD.
5
6
6.16.4 Pro nebulizer unit
Important: The nebulizer unit holds up to
10 ml of liquid medication.
When the nebulizer unit is connected into the
inspiratory tube, the filler cap plug can be
opened and closed in between doses without
causing loss of circuit pressure.
WARNING! Do not attach a continuous
supply of medication to the nebulizer; the
device operates in 5 to 30 minute cycles.
Note: The nebulization rate is >0.2 ml/min,
fill the medication cup and set the
nebulization time accordingly.
6.16.5 Solo nebulizer unit
Important: The nebulizer unit holds up to
6 ml of liquid medication.
When the nebulizer unit is connected into the
inspiratory tube, the filler cap plug can be
opened and closed in between doses without
causing loss of circuit pressure.
WARNING! Do not use the Solo nebulizer
in conjunction with the administration of
volatile anaesthetics as this may have an
adverse effect on the Solo nebulizer or
T piece plastics.
Important: This is a single patient use
device not to be used on more than one
patient to prevent cross-infection.
SERVO-air Ventilator System v2.2, User's Manual
107
|
Ventilatory settings and functions
| 6 |
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...