It is also very important to set the lower and
upper alarm limits for expired minute volume.
The inspiratory rise time should be set to a
comfortable value for the patient. In PS, the
inspiratory rise time should normally be
increased.
The End inspiration setting is important to
patient comfort and ventilator synchronization
with the patient. If the patient's expiratory
resistance is high, the End inspiration setting
should be raised to guarantee enough time for
expiration.
It is important to monitor how this affects the
tidal volume.
68
SERVO-air Ventilator System v2.2, User's Manual
| 5 |
Ventilation modes
|